Infrared Temperature Calibrator
FEATURES
• For calibrating non-contact infrared thermometers up to 500°C (932°F)
• Temperature controlled black-body target with
a diameter of 2.25" (57mm)
• Wide field of view to compensate for optical
variations in individual infrared thermometers
• Simply set the desired temperature in increments of 0.1°, wait for the calibrator to attain
equilibrium, point the infrared thermometer at
the target, and measure the radiated energy
• Contact RTD reference well is located directly
behind the blackbody target to assure even
higher accuracies of ±0.1°C
ORDERING INFORMATION
BX-500 - Infrared Temperature Calibrator
1400C - Hard Carrying Case
SPECIFICATIONS
|
Temperature Range
|
50.0 to 500.0°C (122.0 to 932.0°F)
|
|
Resolution
|
0.1°C (0.1°F)
|
|
Accuracy
|
±0.8°C (±1.6°F): ≤100°C (212°F);
±1.6°C (±3.2°F): 100 to ≤200°C (212 to 392°F);
±2.8°C (±5.6°F): 200 to ≤500°C (392 to ≤932°F)
|
|
Stability
|
±0.1°C (±0.2°F): ≤100°C (212°F);
±0.2°C (±0.4°F): 100 to ≤350°C (212 to 662°F);
±0.4°C (±0.8°F): 350 to ≤500°C (662 to ≤932°F)
|
|
Target Emissivity
|
Fixed at 0.95
|
|
Cooling Time
|
45 minutes to 100°C (122°F)
|
|
Heating Time
|
40 minutes to maximum
|
|
Aperture Diameter
|
57mm
|
|
Power Supply
|
1.25A/110VAC (±10%), 2.5A
|
|
Dimensions
|
180 x 114 x 233mm
|
|
Weight
|
3 kg
|
Product added : July 03 2012 03:26:19 AM

|
 |
|